Formaldehyde Climate Chamber

PRODUCT DETAILS

Product Details

Environmental Conditions

  • Operating Temperature: Generally, the temperature should be between 15°C to 40°C, as required by climate chambers produced by Guangzhou Dejun Instruments.
  • Operating Humidity: Typically, the environmental humidity should range from 5% to 90% RH.

Temperature Control

  • Adjustable Range: Temperature is usually adjustable between 18°C to 30°C or 10°C to 40°C, with a deviation typically between ±1°C to ±0.5°C.
  • Uniformity and Fluctuation: Temperature uniformity is generally around 1.0°C, and fluctuation is between ±0.3°C to ±0.5°C.

Humidity Control

  • Adjustable Range: Relative humidity is typically adjustable between 40% to 60% or 30% to 80%, with a deviation between ±3% RH to ±5% RH.
  • Fluctuation and Uniformity: Humidity fluctuation is generally between ±2% RH to ±3% RH, with uniformity around 3% RH.

Air Circulation

  • Air Velocity: Typically ranges from 0.1m/s to 0.5m/s or 0.1m/s to 2m/s.
  • Air Exchange Rate: Generally ranges from 0.2m³/h to 2.0m³/h or 0.2 times/hour to 3 times/hour, with a deviation of ±3% to ±5%.

Background Concentration and Gas Tightness

  • Background Concentration: The formaldehyde concentration in the air of the chamber should not exceed 0.006mg/m³ when empty.
  • Gas Tightness: Under a 1000Pa overpressure condition, the gas leakage rate should not exceed 0.005m³/min or 10⁻³×1m³/min.

Here’s the translation of the additional specifications:

Volume and Dimensions

  • Effective Volume: The standard effective volume of a formaldehyde testing climate chamber is 1.0±0.01m³.

  • External Dimensions: The external dimensions of the equipment vary depending on the manufacturer, commonly including sizes such as 1800×1150×1850mm, 1480×1140×1700mm, etc.

Other Parameters

  • Noise Level: The noise level during operation of the climate chamber typically does not exceed 60dB.

  • Continuous Operation Time: The continuous operation time is typically no less than 28 days.

  • Power Supply: Common power specifications include AC 220V/50Hz or AC 380V/50Hz.

  • Power Consumption: The starting power is typically between 7kW and 8kW, while the operating power is around 4kW.